Table 1: Typical pulse sequence protocol for liver tumor examination with Gadoxetate disodium-enhanced MRI.

MRI pulse sequence protocol for liver tumor examination
Pulse Sequence Plane TR (ms) TE (ms) Flip angle (°) Field of View (mm) Matrix Voxel Size
2D T1-weighted GRE (out/in phase) (FLASH) Axial 100 2.4/4.8 70 200 256 1.1×0.9×5.0 mm
2D T1-weighted GRE (FLASH) Axial 180 2.3 20 380 320 1.3×0.9×4.0 mm
Injection of gadoxetate disodium (0.025 mmol/kg body weight)
3D T1-weighted FS GRE pre contrast (VIBE)
(Pre contrast/ arterial/ venous /equilibrium)
Axial 5.6 2.6 10 200 256 1.0×0.8×3.0 mm
2D T2-weighted single shot FSE (HASTE) (Phase Partial Fourier 4/8) Axial, Coronal, Axial 1000 86 150 325 256 1.3×1.3×6.0 mm
2D coronal balanced steady-state free precession (TruFISP) Axial 4.8 2.4 68 300 256 1.3×1.2×6.0 mm
2D T2-weighted fat saturated with radial k-space coverage (PROPELLER/BLADE) Axial and Coronal 113 5500 90 220 320 0.7×0.7×4.0 mm
Diffusion-weighted imaging (b= 0, 200, 500, 800, 1000) Axial 4,000 79 180 360 192 1.1×1.1×6.0 mm
3D T1-weighted GRE hepatocyte phase (VIBE) Axial and Coronal 5.6 2.3 10 360 256 1.0×0.8×3.0 mm
Note: FSE = fast spin-echo, GRE = gradient-recalled echo.