Table 4: Reported rates* of manual rescanning.
|
# of Slides |
Total Rescan |
Slide-related Issue |
Scanning/Image Related Issue |
|||
|
Tissue/Label Detection Error |
Image Quality or Incomplete Tissue |
Total |
||||
|
Epredia E1000 Dx Digital Pathology Solution |
1,796 |
5.8% |
5.6% |
0.0% |
0.2% |
0.2% |
|
PIPS UFS |
3,390 |
6.0% |
2.1% |
2.3% |
1.6% |
3.9% |
|
Leica Aperio AT2 DX |
2,119 |
1.8% |
n.a. |
1.0% |
0.8% |
1.8% |
<* Systems for which published data on rescan rates was not publicly available were omitted from the table.