Table 1: Summary statistics of phacoemulsification of 2.0 mm porcine lens cubes at varying bottle height, vacuum, and power settings.

Bottle height (cm) Vacuum (mmHg) Power (%) Efficiency (s) ± SD Chatter ± SD
60 600 60 0.841 ± 0.21 0.05 ± 0.22
80 600 60 0.614 ± 0.14 0.1 ± 0.31
100 600 60 0.512 ± 0.079 0 ± 0
100 400 60 0.897 ± 0.29 0.1 ± 0.31
100 500 60 0.519 ± 0.10 0.05 ± 0.22
100 600 60 0.445 ± 0.080 0.05 ± 0.22
100 600 50 1.27 ± 0.46 0.1 ± 0.31
100 600 60 1.22 ± 0.59 0.05 ± 0.22
100 600 70 0.971 ± 0.25 0 ± 0
100 600 80 0.678 ± 0.21 0 ± 0
100 600 90 0.78 ± 0.37 0 ± 0
100 600 100 0.739 ± 0.36 0 ± 0