Table 1: Summary statistics of phacoemulsification of 2.0 mm porcine lens cubes at varying bottle height, vacuum, and power settings.
Bottle height (cm) | Vacuum (mmHg) | Power (%) | Efficiency (s) ± SD | Chatter ± SD |
---|---|---|---|---|
60 | 600 | 60 | 0.841 ± 0.21 | 0.05 ± 0.22 |
80 | 600 | 60 | 0.614 ± 0.14 | 0.1 ± 0.31 |
100 | 600 | 60 | 0.512 ± 0.079 | 0 ± 0 |
100 | 400 | 60 | 0.897 ± 0.29 | 0.1 ± 0.31 |
100 | 500 | 60 | 0.519 ± 0.10 | 0.05 ± 0.22 |
100 | 600 | 60 | 0.445 ± 0.080 | 0.05 ± 0.22 |
100 | 600 | 50 | 1.27 ± 0.46 | 0.1 ± 0.31 |
100 | 600 | 60 | 1.22 ± 0.59 | 0.05 ± 0.22 |
100 | 600 | 70 | 0.971 ± 0.25 | 0 ± 0 |
100 | 600 | 80 | 0.678 ± 0.21 | 0 ± 0 |
100 | 600 | 90 | 0.78 ± 0.37 | 0 ± 0 |
100 | 600 | 100 | 0.739 ± 0.36 | 0 ± 0 |